SCHEDULE 8CONTAINMENT MEASURES

PART I

1.  In this Schedule—

“GMMs” means genetically modified micro-organisms;

“HEPA” means High Efficiency Particulate Air;

“inactivation” means the complete or partial destruction of GMMs so as to ensure that any contact between the GMMs and humans or the environment is limited to an extent commensurate with the risks identified in the risk assessment and to provide a high level of protection for humans and the environment;

“plant growth facilities” means a structure, whether permanent or impermanent, designed and used principally for growing plants in a controlled and protected environment; and

“risk assessment” means the assessment carried out in accordance with regulation 6.

2.  For the purposes of this Schedule, where, in the final column of Table 1b or 1c, a measure is specified as—

(a)a modification, it shall be read in substitution for the relevant measure in Table 1a;

(b)additional, it shall be read as an addition to the measures in Table 1a, subject to the substitution, where appropriate, of an individual measure in Table 1a by a measure specified as a modification in the Table in question.

3.  For the purposes of this Schedule—

(a)Table 1a describes containment measures applicable to activities involving genetic modification of micro-organisms in laboratories;

(b)Table 1a, read with Table 1b, describes containment measures applicable to activities involving genetic modification of micro-organisms in plant growth facilities;

(c)Table 1a, read with Table 1c, describes containment measures applicable to activities involving genetic modification of micro-organisms in animal units;

(d)Table 2 describes containment measures applicable to activities involving genetic modification of micro-organisms in premises other than those referred to in Tables 1a, 1b and 1c.

PART II

Table 1a:

Containment Measures for Activities Involving Genetic Modification of Micro-organisms in Laboratories

Containment MeasuresContainment Levels
1234

NOTES

1.

In the Table above, “isolation” means, in relation to a laboratory, separation of the laboratory from other areas in the same building, or being in a separate building.

2.

Entry must be through an airlock which is a chamber isolated from the laboratory. The clean side of the airlock must be separated from the restricted side by changing or showering facilities and preferably by interlocking doors.

3.

Where viruses are not retained by the HEPA filters, extra requirements will be necessary for extract air.

4.

Where the autoclave is outside the laboratory in which the activity involving genetic modification of micro-organisms is being undertaken, but within the laboratory suite, there shall be validated procedures for the safe transfer of material into that autoclave, which provide a level of protection equivalent to that which would be achieved by having an autoclave in that laboratory.

1Laboratory suite: isolation (Note 1)not requirednot requiredrequiredrequired
2Laboratory: sealable for fumigationnot requirednot requiredrequiredrequired
Equipment
3Surfaces impervious to water, resistant to acids, alkalis, solvents, disinfectants and decontamination agents and easy to cleanrequired for benchrequired for benchrequired for bench and floorrequired for bench, floor ceiling and walls
4Entry to lab via airlock (Note 2)not requirednot requiredrequired where and to extent the risk assessment shows it is requiredrequired
5Negative pressure relative to the pressure of the immediate surroundingsnot requiredrequired where and to extent the risk assessment shows it is requiredrequiredrequired
6Extract and input air from the laboratory shall be HEPA filterednot requirednot requiredHEPA filters required for extract airHEPA filters required for input and extract air (Note 3)
7Microbiological safety cabinet/enclosurenot requiredrequired where and to extent the risk assessment shows it is requiredrequired, and all procedures with infective materials required to be contained within a cabinet/enclosureClass III cabinet required
8Autoclaverequired on siterequired in the buildingrequired in the laboratory suite (Note 4)double ended autoclave required in laboratory
System of work
9Access restricted to authorised personnel onlynot requiredrequiredrequiredrequired (via airlock key procedure)
10Specific measures to control aerosol disseminationnot requiredrequired so as to minimiserequired so as to preventrequired so as to prevent
11Showernot requirednot requiredrequired where and to extent the risk assessment shows it is requiredrequired
12Protective clothingsuitable protective clothing requiredsuitable protective clothing requiredsuitable protective clothing required; footwear required where and to extent the risk assessment shows it is requiredcomplete change of clothing and footwear required before entry and exit
13Glovesnot requiredrequired where and to extent the risk assessment shows they are requiredrequiredrequired
14Efficient control of disease vectors (eg rodents and insects) which could disseminate GMMsrequired where and to extent the risk assessment shows it is requiredrequiredrequiredrequired
15Specified disinfection procedures in placerequired where and to extent the risk assessment shows they are requiredrequiredrequiredrequired
Waste
16Inactivation of GMMs in effluent from handwashing sinks and showers and similar effluentsnot requirednot requiredrequired where and to extent the risk assessment shows it is requiredrequired
17Inactivation of GMMs in contaminated material and wasterequired by validated meansrequired by validated meansrequired by validated meansrequired by validated means
Other measures
18Laboratory to contain its own equipmentnot requirednot requiredrequired, so far as is reasonably practicablerequired
19An observation window or alternative is to be present so that occupants can be seenrequired where and to extent the risk assessment shows it is requiredrequired where and to extent the risk assessment shows it is requiredrequiredrequired
20Safe storage of GMMsrequired where and to extent the risk assessment shows it is requiredrequiredrequiredsecure storage required
21Written records of staff trainingnot requiredrequired where and to extent the risk assessment shows they are requiredrequiredrequired

Table 1b:

Containment Measures for Activities Involving Genetic Modification of Micro-organisms in Plant Growth Facilities (to be read with Table 1a as indicated in paragraph 3)

Containment MeasuresContainment LevelsAdditional/modification
1234

NOTE

1.

A permanent structure refers to a fixed structure with walls, a roof and a floor. Where the permanent structure is a greenhouse, that structure shall also have a continuous waterproof covering and self-closing lockable outer doors, and be located on a site designed to prevent the entry of surface run-off water.

Building
1Permanent structure (Note 1)required where and to extent the risk assessment shows it is requiredrequiredrequiredrequiredModification
Equipment
2Entry via a separate room with two interlocking doorsnot requiredrequired where and to extent the risk assessment shows it is requiredrequired where and to extent the risk assessment shows it is requiredrequired (via airlock key procedure)Additional
3Control of contaminated run-off waterrequired where and to extent the risk assessment shows it is requiredrequired so as to prevent run-offrequired so as to prevent run-offrequired so as to prevent run-offAdditional
System of work
4Effective control of disease vectors such as insects, rodents and arthropods which could disseminate GMMsrequiredrequiredrequiredrequiredAdditional
5Effective control of pollen, seeds and other plant material which could disseminate GMMsrequired where and to extent the risk assessment shows it is requiredrequired so as to minimise disseminationrequired so as to prevent disseminationrequired so as to prevent disseminationAdditional
6Procedures for transfer of living material between the plant growth facilities, protective structure and laboratory shall control dissemination of GMMsrequired so as to minimise disseminationrequired so as to prevent disseminationrequired so as to prevent disseminationrequired so as to prevent disseminationAdditional

Table 1c:

Containment Measures for Activities Involving Genetic Modification of Micro-organisms in Animal Units (to be read with Table 1a as indicated in paragraph 3)

Containment MeasuresContainment LevelsAdditional/modification
1234

NOTES

1.

In the Table above, “animal unit” means a building, or separate area within a building, containing an animal facility and other areas including changing rooms, showers, autoclaves and food storage areas.

2.

In the Table above and in Note 1 above, “animal facility” means a facility normally used to house stock, breeding or experimental animals or one which is used for the performance of minor surgical procedures on animals.

3.

In the Table above, “isolators” means transparent boxes where small animals are contained within or outside a cage; for large animals, isolated rooms may be more appropriate.

Facilities
1Isolation of animal unit (Note 1)required where and to extent the risk assessment shows it is requiredrequiredrequiredrequiredModification
2Animal facilities (Note 2) separated by lockable doorsrequired where and to extent the risk assessment shows they are requiredrequiredrequiredrequiredAdditional
3Animal facilities (cages, etc) designed to facilitate decontamination (waterproof and easily washable material)required where and to extent the risk assessment shows they are requiredrequired where and to extent the risk assesment shows they are requiredrequiredrequiredAdditional
4Floor, walls and ceiling easily washablerequired where and to extent the risk assessment shows they are requiredrequired for floorrequired for floor and wallsrequired for floor, walls and ceilingModification
5Appropriate filters on isolators or isolated rooms (Note 3)not requiredrequired where and to extent the risk assessment shows they are requiredrequiredrequiredAdditional
6Incinerator for disposal of animal carcassesrequired to be accessiblerequired to be accessiblerequired to be accessiblerequired to be on siteAdditional
7Appropriate barriers at the room exit, and at drains or ventilation duct workrequiredrequiredrequiredrequiredAdditional
8Animals kept in appropriate containment facilities, such as cages, pens, tanks or isolatorsrequired where and to extent the risk assessment shows it is requiredrequired where and to extent the risk assessment shows it is requiredrequired where and to extent the risk assessment shows it is requiredrequired where and to extent the risk assessment shows it is requiredAdditional

Table 2:

Containment Measures for Activities Involving Genetic Modification of Micro-organisms in Premises other than those referred to in Tables 1a, 1b and 1c

Containment MeasuresContainment Levels
1234
General
1Viable micro-organisms shall be contained in a system which separates the process from the workplace and wider environment (closed system)required where and to extent the risk assessment shows it is requiredrequiredrequiredrequired
2Closed systems located within a controlled areanot requiredrequired where and to extent the risk assessment shows they are requiredrequiredrequired and required to be purpose built
3Control of exhaust gases from the closed systemnot requiredrequired so as to minimise releaserequired so as to prevent releaserequired so as to prevent release
4Control of aerosols during sample collection, addition of material to a closed system or transfer of material to another closed systemrequired where and to extent the risk assessment shows it is requiredrequired so as to minimise releaserequired so as to prevent releaserequired so as to prevent release
5Inactivation of bulk culture fluids before removal from the closed systemrequired where and to extent the risk assessment shows it is requiredrequired by validated meansrequired by validated meansrequired by validated means
6Seals shall be designed so as to minimise or prevent releasenot requiredrequired so as to minimise releaserequired so as to prevent releaserequired so as to prevent release
7The controlled area designed to contain spillage of the entire contents of the closed systemrequired where and to extent the risk assessment shows it is requiredrequired where and to extent the risk assessment shows it is requiredrequiredrequired
8The controlled area sealable to permit fumigationnot requiredrequired where and to extent the risk assessment shows it is requiredrequired where and to extent the risk assessment shows it is requiredrequired
9Biohazard signs postedrequired where and to extent the risk assessment shows it is requiredrequiredrequiredrequired
Equipment
10Entry via airlocknot requirednot requiredrequired where and to extent the risk assessment shows it is requiredrequired
11Surfaces resistant to water, acids, alkalis, solvents, disinfectants and decontamination agents and easy to cleanrequired for any benchrequired for any benchrequired for floor and any benchrequired for bench, floor, ceiling and walls
12Specific measures to adequately ventilate the controlled areas in order to minimise air contaminationrequired where and to extent the risk assessment shows they are requiredrequired where and to extent the risk assessment shows they are requiredrequired where and to extent the risk assessment shows they are requiredrequired
13The controlled area maintained at an air pressure negative to the immediate surroundingsnot requirednot requiredrequired where and to extent the risk assessment shows it is requiredrequired
14Extract and input air from the controlled area shall be HEPA filterednot requirednot requiredrequired for extract air, optional for input airrequired for input and extract air
System of work
15Access restricted to authorised personnel onlynot requiredrequiredrequiredrequired
16Decontamination and washing facilities provided for personnelrequiredrequiredrequiredrequired
17Personnel shall shower before leaving the controlled areanot requirednot requiredrequired where and to extent the risk assessment shows it is requiredrequired
18Personnel shall wear protective clothingwork clothing requiredwork clothing requiredrequiredcomplete change required before exit and entry
19Written procedures and records of staff trainingnot requirednot requiredrequiredrequired
Waste
20Inactivation of GMMs in effluent from handwashing sinks and showers or similar effluentsnot requirednot requiredrequired where and to extent the risk assessment shows it is requiredrequired
21Inactivation of GMMs in contaminated material and waste including those in process effluent before final dischargerequired by validated meansrequired by validated meansrequired by validated meansrequired by validated means

Yn ôl i’r brig

Options/Help

Print Options

Close

Mae deddfwriaeth ar gael mewn fersiynau gwahanol:

Y Diweddaraf sydd Ar Gael (diwygiedig):Y fersiwn ddiweddaraf sydd ar gael o’r ddeddfwriaeth yn cynnwys newidiadau a wnaed gan ddeddfwriaeth ddilynol ac wedi eu gweithredu gan ein tîm golygyddol. Gellir gweld y newidiadau nad ydym wedi eu gweithredu i’r testun eto yn yr ardal ‘Newidiadau i Ddeddfwriaeth’.

Gwreiddiol (Fel y’i Deddfwyd neu y’i Gwnaed): Mae'r wreiddiol fersiwn y ddeddfwriaeth fel ag yr oedd pan gafodd ei deddfu neu eu gwneud. Ni wnaed unrhyw newidiadau i’r testun.

Close

Dewisiadau Agor

Dewisiadau gwahanol i agor deddfwriaeth er mwyn gweld rhagor o gynnwys ar y sgrin ar yr un pryd

Close

Rhagor o Adnoddau

Gallwch wneud defnydd o ddogfennau atodol hanfodol a gwybodaeth ar gyfer yr eitem ddeddfwriaeth o’r tab hwn. Yn ddibynnol ar yr eitem ddeddfwriaeth sydd i’w gweld, gallai hyn gynnwys:

  • y PDF print gwreiddiol y fel deddfwyd fersiwn a ddefnyddiwyd am y copi print
  • rhestr o newidiadau a wnaed gan a/neu yn effeithio ar yr eitem hon o ddeddfwriaeth
  • manylion rhoi grym a newid cyffredinol
  • pob fformat o’r holl ddogfennau cysylltiedig
  • slipiau cywiro
  • dolenni i ddeddfwriaeth gysylltiedig ac adnoddau gwybodaeth eraill
Close

Rhagor o Adnoddau

Defnyddiwch y ddewislen hon i agor dogfennau hanfodol sy’n cyd-fynd â’r ddeddfwriaeth a gwybodaeth am yr eitem hon o ddeddfwriaeth. Gan ddibynnu ar yr eitem o ddeddfwriaeth sy’n cael ei gweld gall hyn gynnwys:

  • y PDF print gwreiddiol y fel gwnaed fersiwn a ddefnyddiwyd am y copi print
  • slipiau cywiro

liciwch ‘Gweld Mwy’ neu ddewis ‘Rhagor o Adnoddau’ am wybodaeth ychwanegol gan gynnwys

  • rhestr o newidiadau a wnaed gan a/neu yn effeithio ar yr eitem hon o ddeddfwriaeth
  • manylion rhoi grym a newid cyffredinol
  • pob fformat o’r holl ddogfennau cysylltiedig
  • dolenni i ddeddfwriaeth gysylltiedig ac adnoddau gwybodaeth eraill