After exit day there will be three versions of this legislation to consult for different purposes. The legislation.gov.uk version is the version that applies in the UK. The EU Version currently on EUR-lex is the version that currently applies in the EU i.e you may need this if you operate a business in the EU.
The web archive version is the official version of this legislation item as it stood on exit day before being published to legislation.gov.uk and any subsequent UK changes and effects applied. The web archive also captured associated case law and other language formats from EUR-Lex.
EU Directives are being published on this site to aid cross referencing from UK legislation. After IP completion day (31 December 2020 11pm) no further amendments will be applied to this version.
This method is suitable for the identification and determination of nitromethane at up to about 0,3 % in cosmetic products packed in aerosol dispensers.
The nitromethane content of the sample determined according to this method is expressed in percentage by mass of nitromethane, in the total aerosol dispenser content.
The nitromethane is identified by colour reaction. Nitromethane is determined gas chromatographically after addition of an internal standard.
All reagents should be of analytical purity.
Dissolve 0,1 g of sodium 3,4-dihydro-3,4-dioxonaphthalene-l-sulphonate in water and dilute to 100 ml.
To 1 ml of sample add 10 ml of 4.1.1 and 1 ml of 4.1.2. A violet coloration indicates the presence of nitromethane.
All reagents must be of analytical quality.
Into a tared 25 ml volumetric flask, introduce about 650 mg of chloroform (5.1.1). Accurately reweigh the flask and contents. Make up to 25 ml with 95 % ethanol (5.1.3). Weigh and calculate the percentage by mass of chloroform in this solution.
Make up in a similar manner to the chloroform reference solution but weigh 270 mg of 2,4-dimethylheptane (5.1.2) into the 25 ml volumetric flask.
Into a 100 ml tared transfer bottle, purged or evacuated according to the procedure described in 5.4 of Chapter II of the abovementioned Directive, introduce about 5 ml of either of the internal standard solutions (5.1.5 or 5.1.6). Use a 10 or 20 ml glass syringe, without needle, adapted to the transfer piece following the technique described in paragraph 5 of Chapter II of the above-mentioned Commission Directive. Reweigh to determine the quantity introduced. Using the same technique, transfer into this bottle about 50 g of the contents of the aerosol dispenser sample. Again reweigh to determine the quantity of sample transferred. Mix well.
Inject about 10 μl using the specified microsyringe (5.2.2). Make five injections.
Into a 50 ml volumetric flask, accurately weigh about 500 mg of nitromethane (5.1.4) and either 500 mg of chloroform (5.1.1) or 210 mg of 2,4-dimethylheptane (5.1.2). Make up to volume with 95 % ethanol (5.1.3). Mix well. Place 5 ml of this solution into a 20 mg volumetric flask. Make up to volume with 95 % ethanol (5.1.3).
Inject about 10 μl using the specified microsyringe (5.2.2). Make five injections.
This is in two parts, the first containing didecyl phthalate on Gas Chrom Q as packing, the second having Ucon 50 HB 280X on Gas Chrom Q as packing. The prepared combined column must yield a resolution ‘R’ equal to, or better than, 1,5, where:
let:
=
retention times (in minutes),
=
peak widths at half height (in millimetres),
=
the chart speed (in millimetres per minute).
As examples the following two parts yield the required resolution:
Column A:
Material: stainless steel.
Length: 1,5 m.
Diameter: 3 mm.
Packing: 20 % didecyl phthalate on Gas Chrom Q (100 to 120 mesh).
Column B:
Materia: stainless steel.
Length: 1,5 m.
Diameter: 3 mm.
Packing: 20 % Ucon 50 HB 280X on Gas Chrom Q (100 to 120 mesh).
A suitable sensitivity setting for the electrometer of the flame ionization detector is 8 × l0-10A.
The following have been found suitable:
Injection port: 150 oC,
Detector: 150 oC,
Column: between 50 and 80 oC depending upon individual columns and apparatus.
Carrier gas: nitrogen.
Pressure: 2,1 bar.
Flow: 40 ml/min
Detector supplies: as specified by the makers of the detector.
If ‘n’ represents nitromethane:
let:
=
its response factor,
=
its mass (in grams) in the mixture,
=
its peak area.
If ‘c’ represents the internal standard, chloroform or 2,4-dimathylheptane:
let:
=
its mass (in grams) in the mixture,
=
its peak area,
then:
(kn is a function o2 the apparatus).
If ‘n’ represents nitromethane:
let:
=
its response factor,
=
its peak area.
If ‘c’ represents the internal standard, chloroform or 2,4-dimethylheptane:
let:
=
its mass (in grans) in the mixture,
=
its peak area,
=
the mass (in grams) of the aerosol transferred,
then the % (m/m) nitromethane in the sample is:
For a nitromethane content of about 0,3 % (m/m), the difference between the results of two determinations carried out in parallel on the same sample should not exceed an absolute value of 0,03 % (m/m).
Norm ISO 5725.
The Whole Directive you have selected contains over 200 provisions and might take some time to download. You may also experience some issues with your browser, such as an alert box that a script is taking a long time to run.
Would you like to continue?
The Schedules you have selected contains over 200 provisions and might take some time to download. You may also experience some issues with your browser, such as an alert box that a script is taking a long time to run.
Would you like to continue?
Latest Available (revised):The latest available updated version of the legislation incorporating changes made by subsequent legislation and applied by our editorial team. Changes we have not yet applied to the text, can be found in the ‘Changes to Legislation’ area.
Original (As adopted by EU): The original version of the legislation as it stood when it was first adopted in the EU. No changes have been applied to the text.
Geographical Extent: Indicates the geographical area that this provision applies to. For further information see ‘Frequently Asked Questions’.
Show Timeline of Changes: See how this legislation has or could change over time. Turning this feature on will show extra navigation options to go to these specific points in time. Return to the latest available version by using the controls above in the What Version box.
Access essential accompanying documents and information for this legislation item from this tab. Dependent on the legislation item being viewed this may include:
This timeline shows the different versions taken from EUR-Lex before exit day and during the implementation period as well as any subsequent versions created after the implementation period as a result of changes made by UK legislation.
The dates for the EU versions are taken from the document dates on EUR-Lex and may not always coincide with when the changes came into force for the document.
For any versions created after the implementation period as a result of changes made by UK legislation the date will coincide with the earliest date on which the change (e.g an insertion, a repeal or a substitution) that was applied came into force. For further information see our guide to revised legislation on Understanding Legislation.
Use this menu to access essential accompanying documents and information for this legislation item. Dependent on the legislation item being viewed this may include:
Click 'View More' or select 'More Resources' tab for additional information including: